"I" sampling valve
|
|
The
Ingold®
sampling valve is designed for sampling medium viscosity
products, with particles up to 1.2mm in diameter. The main
product range are used
in have about 0-800cP viscosity. |
|
Sampling |
Sterilization |
 |
 |
|
|
The "I"
valve is used with bioreactors for sampling of wide range of
products. |
|
|
The
"I"
valve comes with a mini try-clamp connections on in outlets.
Valve heads come with diffrent operations and membranes.
|
General information on the Keofitt®
"I"
sampling valve.
|
|
Temperture
range: |
Rubber membrane: |
0 to 130ºC/32 to 266ºF |
|
|
PTFE membrane: |
0 to 150ºC/32 to 302ºF |
|
Pressure range: |
Valve head type H, Q and N: |
0 to 6 bar(g)/0 to 87 psi |
|
Material:
|
Steel parts: |
AISI 304/316L |
|
|
PTFE, Silicone and EPDM: |
FDA approved |
|
Dimensions: |
Process dimension |
In-/outlet dimension |
|
|
 |
 |
|
body
surface: |
Inner
surface Ra <0.2 µm /
8 µinch |
Electro
polished outside |
|
Material
certificate: |
3.1.B |
|
|